ECG Feature Review — Top 10 Clinical Features for Synthetic ECG Evaluation

Purpose: Justify and document the clinical ECG features implemented in ecgen_eval/features/ for evaluating the quality of synthetically generated ECG signals.

For a plain-language, interactive introduction, see Clinical Features Explained.


Selected Features

# Feature Units Clinical Relevance Implemented
1 RR interval (mean, std) ms Heart rate, rhythm regularity, arrhythmia detection ✓ rr_mean_ms, rr_std_ms
2 Heart Rate (HR) bpm Bradycardia / tachycardia ✓ hr_bpm
3 PR interval ms AV conduction; prolonged → 1st-degree AV block ✓ pr_interval_ms
4 QRS duration ms Bundle branch blocks, ventricular conduction delay ✓ qrs_duration_ms
5 QT interval ms Ventricular repolarisation; prolonged → long-QT syndrome ✓ qt_interval_ms
6 QTc (Bazett) ms Rate-corrected QT: QT / √RR(s); arrhythmia risk marker ✓ qtc_bazett_ms
7 ST deviation mV Displacement from isoelectric baseline; ischaemia / MI marker ✓ st_deviation_mv
8 T-wave amplitude mV Hyperkalemia, ischaemia, LVH; affected by repolarisation ✓ t_amplitude_mv
9 P-wave duration ms Atrial enlargement, inter-atrial conduction delay, AF risk ✓ p_wave_duration_ms
10 HRV — SDNN / RMSSD / pNN50 ms / % Autonomic tone; cardiac risk stratification; vagal activity ✓ hrv_sdnn_ms, hrv_rmssd_ms, hrv_pnn50

Feature Descriptions and Clinical Context

1 & 2 — RR Interval and Heart Rate

The RR interval is the time between consecutive R-peaks measured in milliseconds. It directly encodes heart rate (HR = 60,000 / RR_mean_ms in bpm). Variation in RR interval (captured by rr_std_ms) reflects rhythm regularity. In a generative model evaluation context, the distribution of RR intervals and HR should match between real and synthetic ECG: divergence indicates the model has not learned the cardiac rhythm distribution.

References: - AHA/ACC Electrocardiography Guidelines (2016) - PhysioNet — RR interval definitions: https://physionet.org

3 — PR Interval

The PR interval spans from the onset of the P-wave to the onset of the QRS complex, reflecting the time for atrial depolarisation and AV nodal conduction. Normal range: 120–200 ms. Prolongation indicates first-degree AV block; shortening may indicate Wolff-Parkinson-White or junctional rhythms.

4 — QRS Duration

QRS duration spans the full ventricular depolarisation complex. Normal: 70–110 ms. Prolongation (>120 ms) indicates bundle branch block (LBBB, RBBB) or ventricular conduction abnormality. Synthetic ECGs should replicate the QRS width distribution of the training population.

5 & 6 — QT Interval and QTc

The QT interval spans from QRS onset to T-wave end, reflecting total ventricular repolarisation time. It is rate-dependent; QTc (Bazett's formula: QTc = QT / √RR_s) adjusts for heart rate. Normal QTc: <440 ms (men), <460 ms (women). Prolonged QTc is a marker of risk for torsades de pointes.

Reference: Bazett HC (1920). Heart 7:353–370.

7 — ST Deviation

The ST segment (from QRS offset / J-point to T-wave onset) should be at the isoelectric baseline. Elevation (>1 mm) suggests acute myocardial infarction; depression suggests ischaemia or digoxin effect. In synthetic ECG evaluation, the ST deviation distribution should not be pathologically shifted unless the generative model is trained on a population with such pathology.

8 — T-wave Amplitude

The T-wave reflects ventricular repolarisation. Its amplitude is lead-dependent and varies with pathology: peaked T-waves in hyperkalemia or early STEMI; inverted T-waves in ischaemia and LVH. A generative model should reproduce the lead-specific T-wave amplitude distribution.

9 — P-wave Duration

P-wave duration (normal: 80–120 ms) reflects atrial depolarisation time. Prolongation indicates inter-atrial conduction delay (P-mitrale) and is a risk factor for atrial fibrillation. Absence of P-waves in synthetic output would indicate a fundamental failure to model sinus rhythm.

10 — HRV Metrics (SDNN, RMSSD, pNN50)

Time-domain HRV measures quantify the variation of RR intervals: - SDNN: Standard deviation of all normal RR intervals — reflects overall HRV, both sympathetic and parasympathetic. - RMSSD: Root mean square of successive RR differences — dominated by parasympathetic (vagal) modulation; robust to non-stationarity. - pNN50: Percentage of successive RR differences >50 ms — correlated with RMSSD; indicator of vagal tone.

Low HRV (reduced SDNN/RMSSD) is associated with increased cardiac risk. Generative models that produce overly regular synthetic ECGs (low HRV) would fail these checks.

References: - Task Force of ESC/NASPE (1996). Heart rate variability. Circulation 93:1043–1065. - Makowski et al. (2021). NeuroKit2: A Python toolbox for neurophysiological signal processing. Behavior Research Methods.


Algorithm Implementation Choices

Task Primary (neurokit2) Fallback (scipy / NumPy)
R-peak detection nk.ecg_peaks() (Pan-Tompkins) Scipy bandpass + find_peaks
Full delineation (P, QRS, T) nk.ecg_delineate(method="peak") Heuristic windows around R-peaks
HRV computation Manual from RR series Manual from RR series (same)

The heuristic fallback windows are based on typical ECG timing for a 60-bpm sinus rhythm and are adequate for distribution-level comparisons across large cohorts of real vs synthetic ECG. For individual beat-level clinical diagnostics, the neurokit2 path is strongly recommended.


References

  1. Goldberger AL et al. (2000). PhysioBank, PhysioToolkit, PhysioNet. Circulation 101:e215–e220.
  2. Makowski D et al. (2021). NeuroKit2. Behavior Research Methods 53:1689–1696.
  3. Task Force (ESC/NASPE) (1996). Heart rate variability: standards of measurement. Circulation 93:1043–1065.
  4. Bazett HC (1920). An analysis of the time relations of electrocardiograms. Heart 7:353–370.
  5. AHA/ACC/HRS (2016). Recommendations for the standardization and interpretation of the ECG. J Am Coll Cardiol 67:2516–2524.
  6. Pan J, Tompkins WJ (1985). A real-time QRS detection algorithm. IEEE Trans Biomed Eng 32:230–236.